EasyManua.ls Logo

Endress+Hauser Liquisys M CLM253 - Measuring Cables and Sensor Connection

Endress+Hauser Liquisys M CLM253
122 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Liquisys M CLM223/253 Electrical connection
Endress+Hauser 21
31 +
Sensor
32 -
33 +
34 -
85 +
86 -
93 +
94 - 42
81 + 43
82 - 41
Temp.
(opt.)
+15V
10mA
2
1
NC
NC
NC
131081-4E
S
13
12
YE
GN
Mains
Contacts: max. 2A, AC:250V/500VA, DC:30V/60W
L1 +
AC DC
N -
REL1
REL2
REL3
REL4
4..20mA
24
23
49
COM
COM
COM
COM
-
+
59
54
55
56
52
51
53
58
57
48
47
131084-4C
15
11
WH
L1 +
AC DC
N -
1
Lf
84
16
83
Binary
Binary
A0008912
 11 Panel-mounted device connection sticker
1 Ground terminal
5.1.2 Measuring cables and sensor connection
You require shielded special measuring cables to connect conductivity sensors to the
transmitter. The following multi-core, pre-terminated cable types can be used:
Sensor type Cable Extension
Two-electrode sensor with or without
temperature sensor Pt 100
CYK71
CPK9* (for CLS16)
VBM box + CYK71 cable
Inductive sensor CLS50, CLS52 Fixed cable on sensor VBM box + CLK6 cable
* High-temperature version without PML
Maximum cable length
Conductive measurement of
conductivity
Max. 100 m (328 ft) with CYK71
Resistance measurement Max 15 m (49.2 ft) with CYK71
Inductive measurement of conductivity Max. 55 m ( 180 ft) with CLK5 (incl. sensor cable)
Structure and termination of the measuring cables

Table of Contents