Commissioning Liquisys M CLM223/253
46 Endress+Hauser
First enter the desired current output configuration into the following blank table with a
pencil. Calculate the resulting signal distance per mA to observe the necessary minimum
slope. Then enter the values in the device.
Current output 1 Current output 2
Value pair [mS/cm]
[%] [°C]
Current
[mA]
Distance
per mA
[mS/cm]
[%] [°C]
Current
[mA]
Distance
per mA
1
2
3
4
5
6
7
8
9
10
Functions marked in italics are not supported by the basic device version.
Coding Field Range of
adjustment
(factory settings
in bold)
Display Info
O
CURRENT
OUTPUT
function group
O
OUTPUT
A0025026-EN
Configuration of the current
output (does not apply for
PROFIBUS).
O1 Select current
output
Out 1
Out 2
O1
Sel.Out
Out1
A0025027-EN
A characteristic can be
selected for every output.
O2 Select measured
variable for 2nd
current output
°C
mS/cm, MΩ, %
Contr
O2
Parameter
¡C
A0025028-EN
R237=curr (current output 2)
can only be selected if
O2=Contr (controller) is
selected (relay card required).
O3 (1) Enter the
characteristic
type
Lin = linear (1)
Sim = simulation
(2)
Tab = table (3)
O3
Sel.Type
lin
A0025029-EN
The characteristic can have a
positive or negative slope for
the measured value output.
In the case of actuating
variable output (O2 = Contr),
an increasing current
corresponds to an increasing
actuating variable.