Commissioning Liquisys M CLM223/253
48 Endress+Hauser
Coding Field Range of
adjustment
(factory settings
in bold)
Display Info
O333 Select table value
pair
1
1 to no. elem.
Assign
O333
Sel.Elem.
1
A0025044-EN
The system runs through the
O333 to O335 function chain
as often as indicated in O332.
"Assign" appears as the last
step. The display goes to
O336 after confirmation.
O334 Enter x value Cond/ind: 0.00
μS/cm
MOhm:
0.00kΩ*cm
Conc: 0.00 %
Temp: 0.00 °C
O334
µS/cm
Meas.val.
0.00
A0025045-EN
x value = measured value
specified by user.
O335 Enter y value 0.00 mA
0.00 to 20.00 mA
O335
mA
mA value
0.00
A0025048-EN
y value = user-defined current
value pertaining to O334.
Return to O333 until all
values are entered.
O336 Message as to
whether table
status is OK
Yes
No
O336
Status ok
yes
A0025049-EN
Back to O3.
If status = no, correct the table
(all settings made up until
now remain intact) or go back
to the measuring mode (table
is deleted).