EasyManua.ls Logo

Endress+Hauser Micropilot FMR53 - Page 22

Endress+Hauser Micropilot FMR53
106 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Micropilot FMR53, FMR54
22 Endress+Hauser
Terminal assignment 4-wire: 4-20 mA HART (10.4 to 48 V
DC
)
3
1
+
L+
4
2
-
L-
1
3
2
A0036516
 8 Terminal assignment 4-wire: 4-20 mA HART (10.4 to 48 V
DC
)
1 Connection 4-20 mA HART (active): terminals 3 and 4
2 Connection supply voltage: terminals 1 and 2
3 Terminal for cable screen
Block diagram 4-wire: 4-20 mA HART (10.4 to 48 V
DC
)
5
4
1
2
3
+
L+
-
L-
Y
I
6
7
3
4
1
2
A0036526
 9 Block diagram 4-wire: 4-20 mA HART (10.4 to 48 V
DC
)
1 Evaluation unit, e.g. PLC
2 HART communication resistor (≥ 250 Ω); observe maximum load
3 Connection for Commubox FXA195 or FieldXpert SFX350/SFX370 (via VIATOR Bluetooth modem)
4 Analog display device; observe maximum load
5 Cable screen; observe cable specification
6 Measuring device
7 Supply voltage; observe terminal voltage, observe cable specification

Other manuals for Endress+Hauser Micropilot FMR53

Related product manuals