EasyManua.ls Logo

Endress+Hauser Micropilot FMR53 - Storage Temperature; Climate Class

Endress+Hauser Micropilot FMR53
106 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Micropilot FMR53, FMR54
58 Endress+Hauser
FMR54
Planar antenna
GT19 housing (plastic PBT)
Temperature specifications: °C (°F)
T
p
P1
T
a
P2
P3
P4
P5
A0019351
Power supply; output (item 2
in the product structure)
P1 P2 P3 P4 P5 P6
T
p
T
a
T
p
T
a
T
p
T
a
T
p
T
a
T
p
T
a
T
p
T
a
A -40
(-40)
80
(176)
80
(176)
80
(176)
150
(302)
60
(140)
150
(302)
-40
(-40)
-40
(-40)
-40
(-40)
- -
B
Switch output not used
-40
(-40)
76
(169)
76
(169)
76
(169)
150
(302)
60
(140)
150
(302)
-40
(-40)
-40
(-40)
-40
(-40)
- -
B
Switch output used
-40
(-40)
60
(140)
60
(140)
60
(140)
150
(302)
41
(106)
150
(302)
-40
(-40)
-40
(-40)
-40
(-40)
- -
C
Channel 2 not used
-40
(-40)
82
(180)
82
(180)
82
(180)
150
(302)
60
(140)
150
(302)
-40
(-40)
-40
(-40)
-40
(-40)
- -
C
Channel 2 used
-40
(-40)
74
(165)
74
(165)
74
(165)
150
(302)
57
(135)
150
(302)
-40
(-40)
-40
(-40)
-40
(-40)
- -
E, G
Switch output not used
-40
(-40)
79
(174)
79
(174)
79
(174)
150
(302)
60
(140)
150
(302)
-40
(-40)
-40
(-40)
-40
(-40)
- -
E, G
Switch output used
-40
(-40)
63
(145)
63
(145)
63
(145)
150
(302)
44
(111)
150
(302)
-40
(-40)
-40
(-40)
-40
(-40)
- -
FMR54
Planar antenna
Housing GT20 (aluminum coated)
Temperature specifications: °C (°F)
T
p
P1
T
a
P2
P3
P4
P5
A0019351
Power supply; output (item 2
in the product structure)
P1 P2 P3 P4 P5 P6
T
p
T
a
T
p
T
a
T
p
T
a
T
p
T
a
T
p
T
a
T
p
T
a
A -40
(-40)
81
(178)
81
(178)
81
(178)
150
(302)
71
(160)
150
(302)
-40
(-40)
-40
(-40)
-40
(-40)
- -
B
Switch output not used
-40
(-40)
82
(180)
82
(180)
82
(180)
150
(302)
72
(162)
150
(302)
-40
(-40)
-40
(-40)
-40
(-40)
- -
B
Switch output used
-40
(-40)
77
(171)
77
(171)
77
(171)
150
(302)
66
(151)
150
(302)
-40
(-40)
-40
(-40)
-40
(-40)
- -
C
Channel 2 not used
-40
(-40)
82
(180)
82
(180)
82
(180)
150
(302)
72
(162)
150
(302)
-40
(-40)
-40
(-40)
-40
(-40)
- -
C
Channel 2 used
-40
(-40)
79
(174)
79
(174)
79
(174)
150
(302)
68
(154)
150
(302)
-40
(-40)
-40
(-40)
-40
(-40)
- -
E, G
Switch output not used
-40
(-40)
83
(181)
83
(181)
83
(181)
150
(302)
73
(163)
150
(302)
-40
(-40)
-40
(-40)
-40
(-40)
- -
E, G
Switch output used
-40
(-40)
78
(172)
78
(172)
78
(172)
150
(302)
68
(154)
150
(302)
-40
(-40)
-40
(-40)
-40
(-40)
- -
K, L -40
(-40)
77
(171)
77
(171)
77
(171)
150
(302)
67
(153)
150
(302)
-40
(-40)
-40
(-40)
-40
(-40)
- -
Storage temperature
–40 to +80 °C (–40 to +176 °F)
–50 °C (–58 °F) with manufacturer declaration on request
Climate class
DIN EN 60068-2-38 (test Z/AD)

Other manuals for Endress+Hauser Micropilot FMR53

Related product manuals