Internal vessel fittings
A0018944
Avoid the location of internal fittings (limit switches, temperature sensors, struts, vacuum rings,
heating coils, baffles etc.) inside the signal beam. Take the beam angle into account.
Avoiding interference echoes
A0016890
Metal orifice plates, installed at an angle to scatter the radar signals, help prevent interference
echoes.