EasyManua.ls Logo

Endress+Hauser Micropilot FMR60 - Dielectric Constant

Endress+Hauser Micropilot FMR60
72 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFTDRAFT DRAFT DRAFT
Micropilot FMR60
Endress+Hauser V. 4, Rev. 2, 19-03-2018 33
FMR60 flange process connection
p
-40
(-40)
+80
(+176)
+130
(+266)
0
(+32)
T
p
[°C]
([°F])
1
2
[bar] ([psi])
3 (43.5)
-1 (-14.5)
0 (0)
A0032615
 11 FMR60: Permitted range for process temperature and process pressure
1 Process connection: flange PP
2 Process connection: flange 316L
FMR60 flange process connection
Feature 100 "Process connection" Process temperature range Process pressure range
XJG:
UNI flange 3"/DN80/80A, PP
XKG:
UNI flange 4"/DN100/100A, PP
XLG:
UNI flange 6"/DN150/150A, PP
–40 to +80 °C (–40 to +176 °F)
p
rel
= –1 to 3 bar (–14.5 to 43.5 psi)
p
abs
< 4 bar (58 psi)
1)
XJJ:
UNI flange 3"/DN80/80A, 316L
XKJ:
UNI flange4"/DN100/100A, 316L
XLJ:
UNI flange 6"/DN150/150A, 316L
–40 to +130 °C (–40 to +266 °F)
1) The pressure range may be further restricted in the event of a CRN approval
Dielectric constant
For liquids
ε
r
≥ 1.9
Please contact Endress+Hauser for applications with lower dielectric constants than indicated.
For dielectric constants (DC values) of many media commonly used in various industries refer
to:
the Endress+Hauser DC manual (CP01076F)
the Endress+Hauser "DC Values App" (available for Android and iOS)

Table of Contents

Other manuals for Endress+Hauser Micropilot FMR60

Related product manuals