Proline Promass 80 Installation
Endress+Hauser 11
2.3.3 Installing the wall-mount housing
"
Caution!
Make sure that the ambient temperature does not exceed the permitted range.
Always install the wall-mount housing in such a way that the cable entries point downwards.
Mounted directly on the wall
Pipe mounting
A0007542
1. Connection compartment
2. Securing screws M6 (max. ø 6.5 mm (0.25"); screw head
max. ø 10.5 mm (0.4")
3. Housing bores for securing screws
A0007543
"
Caution!
Danger of overheating!
If the device is mounted on a warm pipe, make sure
that the housing temperature does not exceed +60 °C
(+140 °F) which is the maximum temperature permitted.
1
2
3
3
90 (3.54)
35 (1.38)
192 (7.56)
81.5 (3.2)
mm (inch)
mm (inch)
Ø 20…70
(Ø 0.79…2.75)
~ ~ 6.1)155 (