Operating Instructions
Proline Promass E 100
Coriolis flowmeter
Modbus RS485
BA01711D/06/EN/03.21
71506065
2021-01-01
Valid as of version
01.03.zz (Device firmware)

Table of Contents

Other manuals for Endress+Hauser Proline Promass E 100

Questions and Answers

  • S
    shawn40Aug 19, 2025
    How to disable write protection on Endress+Hauser Proline Promass E 100 Measuring Instruments?
    • C
      christina27Aug 19, 2025
      If you have no write access to parameters on your Endress+Hauser Measuring Instruments, it's likely that hardware write protection is enabled. To resolve this, set the write protection switch on the main electronics module to the OFF position.
  • A
    ann90Aug 21, 2025
    How to enable write access to parameters on Endress+Hauser Proline Promass E 100 Measuring Instruments?
    • C
      Christian WilliamsAug 21, 2025
      If you don't have write access to the parameters, it might be because the hardware write protection is enabled. Set the write protection switch on the main electronics module to the OFF position.
  • D
    Danielle SmithAug 23, 2025
    Why does my Endress+Hauser Proline Promass E 100 device measure incorrectly?
    • S
      Steven RodriguezAug 23, 2025
      If your Endress+Hauser Measuring Instruments device measures incorrectly, it could be due to a configuration error or operation outside the application's limits. To resolve this, first check and correct the parameter configuration. Second, observe the limit values specified in the Technical Data.
  • J
    John AdamsAug 26, 2025
    What to do if there is no connection via Modbus RS485 for Endress+Hauser Measuring Instruments?
    • K
      Kevin BrennanAug 26, 2025
      If you can't connect to your Endress+Hauser Measuring Instruments via Modbus RS485, several factors could be at play. First, check the terminal assignment to ensure the Modbus RS485 bus cable is connected correctly. Second, verify the pin assignment of the device plug. Third, inspect the terminating resistor to confirm the Modbus RS485 cable is correctly terminated. Finally, check the Modbus RS485 configuration to ensure the communication interface settings are correct.
  • A
    alynchAug 27, 2025
    Why is the green power LED on the Endress+Hauser Proline Promass E 100 Measuring Instruments transmitter dark?
    • G
      grodriguezAug 27, 2025
      If the green power LED on the main electronics module of your Endress+Hauser Measuring Instruments transmitter is dark, it could be due to two reasons. First, the supply voltage may not match what's specified on the nameplate. Ensure you apply the correct supply voltage. Second, the power supply cable might be connected incorrectly. Verify the terminal assignment.
  • S
    Shane NelsonAug 30, 2025
    Why is the green power LED on Safety Barrier Promass 100 of my Endress+Hauser Measuring Instruments dark?
    • B
      betty51Aug 30, 2025
      If the green power LED on the Safety Barrier Promass 100 of your Endress+Hauser Measuring Instruments is dark, there are a couple of potential causes. One, the supply voltage might not match the specification on the nameplate; ensure you're applying the correct voltage. Two, the power supply cable could be connected incorrectly; double-check the terminal assignment.
  • C
    Christopher JohnsonSep 1, 2025
    Why is my Endress+Hauser Measuring Instrument measuring incorrectly?
    • T
      Timothy SmithSep 1, 2025
      Incorrect measurements can occur if there's a configuration error or if the device is being used outside of its intended application. To resolve this: 1. Check and correct the parameter configuration. 2. Make sure to observe the limit values specified in the "Technical Data".
  • K
    kelly87Sep 5, 2025
    How to fix no or incomplete display of contents in the Web browser for Endress+Hauser Proline Promass E 100?
    • A
      Amy SmithSep 5, 2025
      If the contents in the web browser are not displayed correctly, the issue might be that JavaScript is not enabled. To resolve this: 1. Enable JavaScript. 2. Enter http://XXX.XXX.X.XXX/ basic.html as the IP address.
  • E
    Elizabeth JamesSep 7, 2025
    What to do if Web browser frozen and operation no longer possible with Endress+Hauser Measuring Instruments?
    • S
      srileySep 7, 2025
      If the web browser freezes and operation is no longer possible, there could be a couple of reasons: * **Data transfer active:** Wait until the data transfer or current action is finished. * **Connection lost:** 1. Check the cable connection and power supply. 2. Refresh the Web browser and restart if necessary.
  • S
    Stephanie PalmerSep 9, 2025
    Why Endress+Hauser Measuring Instrument is not connecting to Web server?
    • C
      Cody WilkinsonSep 9, 2025
      If you are unable to connect to the web server, it could be due to the web server being disabled. Via the "FieldCare" operating tool check whether the Web server of the measuring device is enabled and enable it if necessary.

Summary

Safety Information

Personnel Qualifications and Safety

Specifies qualifications and authorization required for personnel performing installation, commissioning, and maintenance.

Intended Use and Media Handling

Details the intended application for liquids and gases and precautions for media compatibility.

Device Installation

Installation Site Conditions

Covers mounting position, environmental factors, and process requirements for installation.

Mounting the Measuring Device

Details the steps for mounting the device, including required tools and preparation.

Post-Installation Inspection

Ensures the device is correctly installed and undamaged through visual inspection.

Electrical Connection and Wiring

Electrical Safety and Cable Specifications

Covers electrical safety standards, cable requirements, and connection conditions.

Connecting Device Components

Guides on connecting the transmitter, safety barrier, and ensuring potential equalization.

Post-Connection Electrical Checks

Verifies the electrical connection, cable integrity, and proper sealing after installation.

Device Operation and Configuration

System Integration and Communication

Modbus RS485 Protocol Information

Details function codes, register information, data types, and the Modbus data map.

Device Commissioning

Pre-Commissioning Steps

Covers function checks and establishing communication via FieldCare before configuration.

Device Parameter Configuration

Guides on setting system units, medium properties, communication interfaces, and advanced settings.

Device Operation

Troubleshooting and Diagnostics

Device Repair and Servicing

Technical Specifications

Endress+Hauser Proline Promass E 100 Specifications

General IconGeneral
BrandEndress+Hauser
ModelProline Promass E 100
CategoryMeasuring Instruments