EasyManua.ls Logo

Endress+Hauser Proline Promass E 100 - Overview of Diagnostic Information

Endress+Hauser Proline Promass E 100
108 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Proline Promass E 100 Modbus RS485 Diagnostics and troubleshooting
Endress+Hauser 71
12.6 Overview of diagnostic information
The amount of diagnostic information increases if the measuring device has one or
more application packages.
Diagnostics for the sensor
Diagnosti
c number
Short text Remedial measures
Status
signal
from the
factory
Diagnostic
behavior
from the factory
022 Sensor temperature 1. Change main electronic module.
2. Change sensor.
F Alarm
044 Sensor drift 1. Check or change main
electronics.
2. Change sensor.
S Alarm
*
046 Sensor limit 1. Inspect sensor.
2. Check process conditions.
S Alarm
*
062 Sensor connection 1. Change main electronic module.
2. Change sensor.
F Alarm
082 Data storage 1. Change main electronic module.
2. Change sensor.
F Alarm
083 Memory content 1. Restart device.
2. Restore S-DAT data.
3. Change sensor.
F Alarm
*
Diagnostic behavior can be changed: "Adapting the diagnostic behavior" section →  70
Diagnostics for the electronics
Diagnosti
c number
Short text Remedial measures
Status
signal
from the
factory
Diagnostic
behavior
from the factory
242 Software incompatible 1. Check software.
2. Flash or change main electronic
module.
F Alarm
261 Electronic modules 1. Restart device.
2. Check electronic modules.
3. Change I/O module or main
electronics.
F Alarm
270 Main electronic failure Change main electronic module. F Alarm
271 Main electronic failure 1. Restart device.
2. Change main electronic module.
F Alarm
272 Main electronic failure 1. Restart device.
2. Contact service.
F Alarm
273 Main electronic failure Replace electronics. F Alarm
274 Main electronic failure Replace electronics. S Warning
*
311 Electronic failure 1. Transfer data or reset device.
2. Contact service.
F Alarm
*
Diagnostic behavior can be changed: "Adapting the diagnostic behavior" section →  70