12.6 Overview of diagnostic information
The amount of diagnostic information increases if the measuring device has one or
more application packages.
Diagnostics for the sensor
Diagnosti
c number
Short text Remedial measures
Status
signal
from the
factory
Diagnostic
behavior
from the factory
022 Sensor temperature 1. Change main electronic module.
2. Change sensor.
F Alarm
044 Sensor drift 1. Check or change main
electronics.
2. Change sensor.
S Alarm
*
046 Sensor limit 1. Inspect sensor.
2. Check process conditions.
S Alarm
*
062 Sensor connection 1. Change main electronic module.
2. Change sensor.
F Alarm
082 Data storage 1. Change main electronic module.
2. Change sensor.
F Alarm
083 Memory content 1. Restart device.
2. Restore S-DAT data.
3. Change sensor.
F Alarm
*
Diagnostic behavior can be changed: "Adapting the diagnostic behavior" section →  70
Diagnostics for the electronics
Diagnosti
c number
Short text Remedial measures
Status
signal
from the
factory
Diagnostic
behavior
from the factory
242 Software incompatible 1. Check software.
2. Flash or change main electronic
module.
F Alarm
261 Electronic modules 1. Restart device.
2. Check electronic modules.
3. Change I/O module or main
electronics.
F Alarm
270 Main electronic failure Change main electronic module. F Alarm
271 Main electronic failure 1. Restart device.
2. Change main electronic module.
F Alarm
272 Main electronic failure 1. Restart device.
2. Contact service.
F Alarm
273 Main electronic failure Replace electronics. F Alarm
274 Main electronic failure Replace electronics. S Warning
*
311 Electronic failure 1. Transfer data or reset device.
2. Contact service.
F Alarm
*
Diagnostic behavior can be changed: "Adapting the diagnostic behavior" section →  70