Proline Promass E 100 Modbus RS485 Technical data
Endress+Hauser 91
Example for maximum measured error
E [%]
Q [%]
0 5 10 15 20 25 30 35 40 45 50 60 6555
0
0.5
1.0
1.5
2.0
2.5
70 75 80 85 90 95 100
A0030289
E Maximum measured error in % o.r. (example)
Q Flow rate in % of maximum full scale value
16.7 Installation
Installation conditions →  18
16.8 Environment
Ambient temperature
range
→  20→  20
Temperature tables
Observe the interdependencies between the permitted ambient and fluid
temperatures when operating the device in hazardous areas.
For detailed information on the temperature tables, see the separate document
entitled "Safety Instructions" (XA) for the device.
Storage temperature –40 to +80 °C (–40 to +176 °F), preferably at +20 °C (+68 °F)
Climate class DIN EN 60068-2-38 (test Z/AD)
Degree of protection Transmitter and sensor
As standard: IP66/67, type 4X enclosure
With the order code for "Sensor options", option CM: IP69 can also be ordered
When housing is open: IP20, type 1 enclosure
Display module: IP20, type 1 enclosure
Safety Barrier Promass 100
IP20
Vibration resistance Oscillation, sinusoidal, following IEC 60068-2-6
2 to 8.4 Hz, 3.5 mm peak
8.4 to 2 000 Hz, 1 g peak
Oscillation, broadband noise following IEC 60068-2-64
10 to 200 Hz, 0.003 g
2
/Hz
200 to 2 000 Hz, 0.001 g
2
/Hz
Total: 1.54 g rms