Repairs
Repair means restoring functional integrity by replacing defective components. Components of
the same type must be used for this purpose. We recommend documenting the repair. This
includes specifying the device serial number, the repair date, the type of repair and the
individual who performed the repair.
The following components may be replaced by the customer's technical staff if genuine spare parts
are used and the appropriate installation instructions are followed:
Component Checking the device after repair
I/O module
Mainboard
Front plane assembly, labelled
Visual inspection to check whether all parts are present and properly
mounted.
Proof test; test sequence A or B.
Cover, aluminum, sight glass
Cover clamp
O-ring, housing
Visual inspection to check whether all parts are present and properly
mounted.
Check the measurement at an arbitrary level.
Electronic box, complete Visual inspection to check whether all parts are present and properly
mounted.
Proof test; test sequence A or B.
Housing filters Visual inspection to check whether all parts are present and properly
mounted.
SD card with holder Visual inspection to check whether all parts are present and properly
mounted.
Proof test; test sequence A or B.
Display set
Display holder, fixing ring
Visual inspection to check whether all parts are present and properly
mounted.
Terminal set, push-in
Terminal set, screw type
Visual inspection to check whether all parts are present and properly
mounted.
Detector unit, servo Visual inspection to check whether all parts are present and properly
mounted.
Calibration must be carried out again (see Operating Instructions).
Repeat the device commissioning check (Diagnostics → Device check).
Mainboard cable to the
electronics, servo
Visual inspection to check whether all parts are present and properly
mounted.
Proof check; test sequence A or B
Sensor module, servo Visual inspection to check whether all parts are present and properly
mounted.
Calibration must be carried out again (see Operating Instructions).
Repeat the device commissioning check (Diagnostics → Device check).
Measuring drum bracket Visual inspection to check whether all parts are present and properly
mounted.
When removing the measuring drum: Calibration must be carried out again
(see Operating Instructions).
Repeat the device commissioning check (Diagnostics → Device check).
Bearing, measuring drum
bracket, PTFE
Visual inspection to check whether all parts are present and properly
mounted.
Calibration must be carried out again (see Operating Instructions).
Repeat the device commissioning check (Diagnostics → Device check).
Measuring wire and or wire
ring
Visual inspection to check whether all parts are present and properly
mounted.
Calibration must be carried out again (see Operating Instructions).
Repeat the device commissioning check (Diagnostics → Device check).
Cover, drum housing Visual inspection to check whether all parts are present and properly
mounted.
Check the measurement at an arbitrary level.
Cover, calibration window/
blind window cover
Visual inspection to check whether all parts are present and properly
mounted.
Check the measurement at an arbitrary level.

Other manuals for Endress+Hauser Proservo NMS80

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Endress+Hauser Proservo NMS80 and is the answer not in the manual?

Endress+Hauser Proservo NMS80 Specifications

General IconGeneral
BrandEndress+Hauser
ModelProservo NMS80
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals