EasyManuals Logo

Endress+Hauser prosonic flow 92 Technical Information

Endress+Hauser prosonic flow 92
16 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #15 background imageLoading...
Page #15 background image
Prosonic Flow 92
Endress+Hauser 15
Materials Transmitter housing Prosonic Flow 92
Handheld housing: plastic case
Flow measuring sensors W
Sensor housing, nickel plated brass: 2.0401 (DIN 17660), C38500 (UNS)
Sensor fixing brackets: 1.4301 (DIN 17440), 304 (AISI)
Sensor contact surface: chemical resistant plastic
Tensioning bands: textile
Flow measuring sensors U
Sensor housing: plastic
Sensor mounting rail: aluminum alloy/plastic
Sensor contact surface: chemical resistant plastic
Tensioning bands: textile
Sensor cables
PVC
Human interface
Display elements Liquid crystal graphic display
240 x 320 dot (with backlight)
Custom configurations for presenting different measured values and status variables
Supported language for display: English, German, French, Italian and Spanish
Operating elements 9 push buttons:
ON, OFF,
,
,
,
, ESC, ENT, LIGHT
Serial communication RS-232C (not isolated)
Transmission speed: max. 9600 BPS
Max. cable length 15 m
Logging functions:
Site data (name, piping, liquid, sensor mounting method, type of sensor) for up to 20 sites
A maximum of 40000 data points (time, velocity, flow rate, totals, analog input, status) can be
stored in the memory.
Certificates and approvals
CE mark The measuring system is in conformity with the statutory requirements of the EC Directives.
Endress+Hauser confirms successful testing of the device by affixing to it the CE mark.
Other standards and
guidelines
EN 60529:
Degrees of protection by housing (IP code)
EN 61010:
Protection Measures for Electrical Equipment for Measurement, Control, Regulation and Labora-
tory Procedures.
EN 61326/A1 (IEC 1326) Emmision to class A requirements:
Electromagnetic compatibility (EMC requirements)
The power adapter is approved according to UL/UL-C and IEC 950.

Other manuals for Endress+Hauser prosonic flow 92

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Endress+Hauser prosonic flow 92 and is the answer not in the manual?

Endress+Hauser prosonic flow 92 Specifications

General IconGeneral
BrandEndress+Hauser
Modelprosonic flow 92
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals