Prosonic M - FOUNDATION Fieldbus Commissioning
Endress+Hauser 51
Function "start mapping" (053)
You have the following options for this function:
• off: Nothing is suppressed.
• on: Starts suppression.
!
Note!
If a mapping already exists, it will be overwritten up to the distance specified in the "range of
mapping" (052) function. Beyond this distance the existing mapping remains unchanged.
Function dist./measured value (008)
After suppression, the measured distance D from the sensor membrane to the product surface is
displayed together with the level. Check that the values correspond to the actual level and/or the
actual distance.
The following cases may occur:
• Distance correct – Level correct -> End of basic calibration
• Distance incorrect – Level incorrect -> An additional interference echo suppression must be
carried out. Go back to the "check distance" (051) function.
• Distance correct – Level incorrect -> Check the value of the "empty calibr." (005) function.
Return to group selection
After the mapping has been recorded the basic calibration is completed and the device automatically
returns to the group selection.