Troubleshooting Prosonic M - FOUNDATION Fieldbus
62 Endress+Hauser
7.1.5 Influence of the error codes on the output signal
The following table describes the influence of the error codes on the status of the cyclic output
values as well as on the parameters BLOCK_ERR and XD_ERROR in the Sensor Block.
The output values are linked to the following measuring values:
Primary Value (PV): level/volume
Secondary Value (SV): distance between sensor membrane and surface of the material measured
Third Value (TV): sensor temperature
W691 Filling noise detected, level ramp is active
Code Error description Action
Code PV Status
SV Status
PV Substatus
SV Substatus
TV Status TV Substatus BLOCK_ER XD_ERROR
A102 BAD Device Failure BAD Device Failure Memory Failure/
Device needs maintenance now
Electronic Failure
W103 Uncertain Non specific GOOD Non specific Other Unspecified Err
A106 BAD Device Failure BAD Device Failure Other Unspecified Err
A110 BAD Device Failure BAD Device Failure Memory Failure/
Device needs maintenance
Electronic Failure
A111 BAD Device Failure BAD Device Failure Memory Failure/
Device needs maintenance now
Electronic Failure
A113 BAD Device Failure BAD Device Failure Memory Failure/
Device needs maintenance now
Electronic Failure
A114 BAD Device Failure BAD Device Failure Memory Failure/
Device needs maintenance now
Electronic Failure
A115 BAD Device Failure BAD Device Failure Device needs maintenance now Unspecified Err
A116 BAD Device Failure BAD Device Failure Device needs maintenance now Unspecified Err
A121 BAD Device Failure BAD Device Failure Memory Failure/
Device needs maintenance now
Electronic Failure
A125 BAD Device Failure BAD Device Failure Memory Failure/
Device needs maintenance now
Electronic Failure
W153 Uncertain Non specific GOOD Non specific Power up No Error
A155 BAD Device Failure BAD Device Failure Device needs maintenace now Electronic Failure
A160 BAD Device Failure BAD Device Failure Memory Failure/
Device needs maintenance now
Electronic Failure
A164 BAD Device Failure BAD Device Failure Memory Failure/
Device needs maintenance now
Electronic Failure
A171 BAD Device Failure BAD Device Failure Memory Failure/
Device needs maintenance now
Electronic Failure
A231 BAD Device Failure BAD Device Failure Device needs maintenance now Unspecified Err
E281
(Warning)
Uncertain Device Failure BAD Device FAilure Device needs maintenance now Electronic Failure
E281
(Alarm)
BAD Device Failure BAD Device Failure Device needs maintenance now Unspecified Err
A502 BAD Device FAilure BAD Device Failure Device needs maintenance now Unspecified Err
A512 Uncertain Non specific GOOD Non specific Other Unspecified Err
A521 BAD Device FAilure BAD Device Failure Device needs maintenance now Unspecified Err
W601 Uncertain configuration error GOOD Non specific Other Configuration Error
W611 Uncertain configuration error GOOD Non specific Other Configuration Error
W621 Uncertain Non specific GOOD Non specific simulation active No Error