21Endress+Hauser
Prosonic T - FMU 130, 131, 230, 231, 232
V0H3 Applications
0: Liquid
including automatic stirrer
suppression
1: Rapid level change
The height changes quickly.
2: Liquid / dome cover including
automatic stirrer suppression.
The instrument is mounted
under a dome cover.
The maximum initial echo
factor is entered as standard.
3. Coarse bulk solids
(grain size from 0.16² / 4 mm)
4. Conveyor belt
Calibration
Reset
A reset causes most of the instrument settings to return to the factory settings. The
following parameters are not affected by a reset:
all linearization parameters (V2H0 – V2H3)
the Tag number (VAH0)
the m/ft selection (V8H2)
Note V8H2 Units of Length
Units of length remain unchanged after carrying out a reset.
They may only be entered directly after a reset. If the units of length are changed at
a later date, then all subsequent entries must be repeated.
Display:
V0H0: Level in %
V0H8: Distance in meters/feet
V0H9: Level in meters/feet
Caution!
All entries which follow (linearization, current output, fixed target suppression) must be
in the same units as those of the calibration.
# VH Entry Text
1 V9H5 333 H Reset instrument
2 V8H2 (0 to 1) H Units of length
0: meter
1: feet
3 V0H1 E (m/ft) H Empty calibration
4 V0H2 F (m/ft) H Full calibration
5 V0H3 H Application
Full
100%
Empty
0%