6. CHECK POINT & CHECK METHOD
【Check 4】 Resistance of Thermistor. (Fig.4)
<EVA.Thermistor resistance measurement>
(at25℃ ~ -16℃)
Test result
2 kΩ - 10 kΩ ※1
0Ω
∞Ω
※1 The resistance increases when the thermistor body is cold.
<Internal thermistor resistance measurement> (Combi type only)
(at25℃ ~ -16℃)
Test result
2 kΩ - 10 kΩ ※1
0Ω
∞Ω
※1 The resistance increases when the thermistor body is cold.
【Check 5】 Output DC voltage measurement from control assembly to power supply
Power supply used Test result Judge
DC12V DC11-14V
DC24V DC23-26V
AC100V
【Check 6】 Output DC voltage measurement from power supply to control assembly
Power supply used Test result Judge
DC12V DC11-14V
DC24V DC23-26V
AC100V
Thermistor disconnection (error display E-09)
Remove the power supply 6P coupler and
measure the DC voltage between the black (-)
wiring section and the red (+) wiring section on
the power supply side 6P coupler section (Fig.
7)
Remove the EVA / thermistor coupler (white) and measure the
resistance between the terminals. (Fig4)
Thermistor short (Error display E-05)
Thermistor disconnection (error display E-06)
DC voltage is measured between the black (-) terminal and
the orange (+) terminal of the power supply 6P coupler (Fig.
6)
Remove the coupler (black) of the internal fan thermistor
and measure the resistance between the terminals. (Fig5)
Thermistor short (Error display E-08)
Orange(+)Black(-)
Black
(-)
Red
(+)