Omniguard™ 4 Owner’s Manual
Page 21
Engineering Solutions Inc
Zero Calibration
The Omniguard 4 does not require calibration between jobs. Internal
temperature compensation and other circuity provides unsurpassed
accuracy over a 30°- 130° Fahrenheit range.
Note: The clicking noise you hear the unit making is
it constantly compensating for zero offset drift. We
do recommend annual calibration verification.
To manually reset the zero point --
1. Disconnect the pressure tubing from INLET 1 and INLET 2.
2. Press MENU to view the Main Menu.
3. Highlight ZERO CALIBRATE using / and press SELECT to
choose zero calibration.
4. Press SAVE to calibrate and return to the Main Menu.
Pressing any other key will return you to the
Monitor Screen without resetting zero.
When completed, the screen will display DONE. After a few moments
the unit will automatically return to the Monitor Screen.
Note: Allow the unit to warm up for 15 minutes
prior to zero calibrating. This permits the pressure
sensor to stabilize its readings. Zero calibration
can only be done when the pressure tubing is not
attached to the unit.
Using Help
To access Help information on a highlighted topic --
Press HELP. Detailed information about the unit’s current status
or the highlighted function will be displayed.
To access the Help Menu --
1. Press HELP twice.
2. Highlight the desired topic using /.
3. Press SELECT to view the information available for that topic.
USB Ports
The Omniguard 4 now has two USB ports. The top port, labeled USB
supports the following devices:
• USB Flash Drive
• Remote Auxiliary Alarm (requires a USB Port Adapter)
• WiFi Adapter (requires a USB Port Adapter)
• Telephone Autodialer (requires a USB Port Adapter)
• Cellular Wireless Modem (requires a USB Port Adapter)
• Telephone Serial Modem (requires a USB Port Adapter)
Pressure Units
The pressure units used by the Omniguard 4 to display and record
pressure readings are selectable.
Pressure Units: (default is Inches WC)
choices are Inches WC (“WC), Millimeters WC (mmWC) or
Pascals (Pa)
To select the desired pressure units --
1. Press MENU to view the Main Menu.
2. Highlight PRESSURE UNITS by using / and press SELECT
to change the setting.
3. Use / to change the pressure units to the desired
measurement.
4. Press SAVE to save the updated settings and return to the Main
Menu.
Pressing the EXIT key will return you to the Main Menu without
saving the updated settings.
Passcode Protect
The passcode protect feature allows the site manager to assign a
passcode that prevents unauthorized changes to the unit’s settings.
Once enabled, the four digit code must be entered to change any
setting. Settings may be viewed without the passcode.
Passcode Protect: (default is OFF)
Passcode: (default passcode is 0 0 0 0)
passcode can be set to ON or OFF
master passcode is 7 7 7 7
To set and enable the Passcode --
1. Press MENU to view the Main Menu.
2. Highlight PASSCODE PROTECTION from the Main Menu using
/ and press SELECT to view or change the setting.
3. Press / to select the passcode On or Off.
5. Press 3/4 to move to the Passcode digits.
4. Set each digit of the passcode using /.
5. Press 3/4 to move from one digit to the next.
6. Press SAVE to save the updated settings and return to the Main
Menu.
Pressing the EXIT key will return you to the Main Menu without
saving the updated settings.
Note: In the event that the passcode is lost or
forgotten it can be reset and settings can be
changed using the master passcode.