The Omniguard 4 provides a report of all the unit’s current settings
(Alarm Setpoints, Log Rates, etc.) and the current system information
(memory status, unit serial number, etc.). This report can be viewed
and printed at any time.
To view or print the Conguration Report --
1. Press MENU to view the Main Menu.
2. Highlight CONFIGURATION REPORT using ▲/▼ and press
SELECT.
3. The Configuration Report will appear in its entirety on the screen
for viewing. Press the PRINT JOB key and the Configuration
Report will begin printing.
Cancel report printing by pressing any key.
Print/Log Rates
The printing and logging of pressure readings begins once the unit
reaches normal operation (status bar indicates NORMAL). There are
two user-definable rates at which printing/logging occurs.
Normal Rate: (default is 15 min.)
• setting choices are 5, 15, 30 minutes or OFF.
• rate at which the highest & lowest pressure readings are
printed and logged to memory during normal operation
• when set to OFF, unit will only log & print alarm readings
Alarm Rate: (default is 15 sec.)
• setting choices are 15, 30, 60 or 120 seconds.
• rate at which pressure readings are printed and logged to
memory during an alarm condition
To change the Print/Log Rate --
1. Press MENU to view the Main Menu.
2. Highlight PRINT/LOG RATES using ▲/▼ and press SELECT to
change settings.
3. Use ▲/▼ to change the NORMAL RATE.
4. Press 3/4 to move from NORMAL RATE to ALARM RATE
and repeat step 3 to change the ALARM RATE setting.
5. Press SAVE to save the updated settings and return to the Main
Menu.
Pressing the EXIT key will return you to the Main Menu without
saving the updated settings.
To conserve paper and log space, the Alarm Print/Log rate will
automatically change to 15 minutes if the unit stays in an alarm
condition for more than 10 minutes. The pressure monitoring rate is
not affected. The Alarm Print/Log rate will return to its original setting
when the next alarm occurs.
Inlet Pressure Response Rate
This setting determines how quickly the unit will react to pressure
changes. Reduce this setting if you are using the unit in conditions
where air pressure fluctuates rapidly and can cause false alarms,
such as in high wind.
Response Rate: (default rate is Medium)
• rate choices are Slow, Medium and Fast
• in windy conditions, set to Slow
• if quick response to small pressure changes is needed, set
to Fast
To change the Response Rate --
1. Press MENU to view the Main Menu.
2. Highlight RESPONSE RATE using ▲/▼ and press SELECT to
view or change the setting.
3. Use ▲/▼ to change the setting.
4. Press SAVE to save the updated setting and return to the Main
Menu.
Pressing the EXIT key will return you to the Main Menu without
saving the updated settings.
Display Contrast & Backlight
The contrast level and backlight setting on the LCD screen allows
the user to adjust the display for optimal viewing in varied lighting
conditions. The backlight automatically dims after 4 minutes to extend
battery and backlight life. It can be reactivated by pressing any key.
Contrast: (default is 50%)
• range is 0% (min) to 100% (max)
Backlight: (default is ON)
• choices are ON or OFF
To change the Contrast or Backlight --
1. Press MENU to view the Main Menu.
2. Highlight DISPLAY from the Main Menu using ▲/▼ and press
SELECT to change the settings.
3. Press 3/4 to choose CONTRAST or BACKLIGHT and use
▲/▼ to change the settings for each.
4. Press SAVE to save the updated settings and return to the Main
Menu.
Pressing the EXIT key will return you to the Main Menu without
saving the updated settings.