UNDERSTANDING PEEP MODE
1. The ADS 2000 has a built in PEEP (Positive End Expiratory
Pressure) mode. To activate the PEEP MODE perform the following:
a. Press the Assist UP button until the unit displays Entering the PEEP MODE
and the alarm beeps.
b. As shown below, the LCD display will now show a PEEP value instead of an
Assist number
NOTE: An " * " (asterisk) is displayed as an indication that the ADS 2000 is in the
PEEP MODE.
NOTE: The Assist MODE will not function when the ADS 2000 is in PEEP MODE.
NOTE: PEEP pressures range from 0 to 9 cm. of H
2
O. Adjust the PEEP
pressure by using the ASSIST UP and ASSIST DOWN buttons.
c. To exit the PEEP MODE Press the ASSIST DOWN button until the
unit displaysEntering the ASSIST MODE” and the alarm beeps.
20 Engler Engineering Corporation ADS 2000