EasyManua.ls Logo

Engler ADS 2000 - Using the ADS 2000 in LAB MODE; LAB MODE Operation Guide

Engler ADS 2000
37 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
USING THE ADS 2000 in LAB MODE
In order to facilitate the ventilation of very small patients the ADS 2000 has a low
pressure LAB MODE. This mode does not have any preset default parameters by
weight; therefore it is advised that you have experience operating the ADS 2000
before using this mode. To enter this mode perform the following:
1. Turn ON the ADS 2000 as usual and allow it to go through the self-test
procedure. The LCD should display the initial default of 20 Kilograms.
NOTE: The start-up self-test may show an error if you attempt to start the ADS 2000
with the input pressure set at 5 PSI, therefore, always start the ADS 2000 in the
NORMAL MODE at 50 PSI then set the 50 PSI / 5 PSI switch to the 5 PSI setting.
2. Now press the WEIGHT DOWN button until the LCD displays looks like the
example below.
3. On the back panel of the ADS 2000 you will find a TOGGLE SWITCH to select the
pressure - 50 PSI for normal operation (and to start up self test) and 5 PSI for LAB
MODE. Set switch to 5 PSI.
This A.D.S has an
internal regulator for
LAB MODE.
Always begin operation
with an input pressure of
50 PSI
50 PSI / 5 PSI
4. In LAB MODE, the Minute Volume per Kilogram is displayed. Tidal Volume is
available anytime when pressing WEIGHT DOWN while SET / RUN switch is on
RUN.
ADS 2000 Engler Engineering Corporation 25