Lab Mode cont.
5. In LAB MODE the FLOW RATES are adjustable between 0.2 to 6.0 Liters Per
Minute.
6. The Breaths Per Minute in LAB MODE are adjustable between 1 and 95.
NOTE: The Breaths Per Minute are in increments of
0.5 for 1 - 12 B.P.M.
1.0 for 13 - 50 B.P.M
2.0 for 50 - 70 B.P.M
5.0 for 70 - 95 B.P.M
7. The MASK function in LAB MODE does not require the use of a Mask Adapter.
USE OF EXTERNAL EQUIPMENT WITH THE ADS 2000
Connecting any external apparatus to the ADS 2000 may adversely affect the
operation of the unit. EEC will not be held liable for any damage to the unit, injury
or death to the patient if any non-approved equipment is used.
Always test for correct operation on the Test Lung prior to using the ADS 2000 on a
patient.
26 Engler Engineering Corporation ADS 2000