Troubleshooting cont.
If the Battery Low LED stays on after the unit has been powered you may have one of
the following problems:
a. There is no power in the wall socket that you are connecting the unit to.
b. The power adapter is damaged.
c. The battery has expired and needs to be replaced. To have the battery replaced,
send the unit to Engler Engineering Corporation. See engler411.com for return,
repair, and loaner forms, click the “repair” tab..
d. The female power socket in the back of the ADS 2000 has been damaged. See
engler411.com for return, repair, and loaner forms, click the “repair” tab.
6. Make sure that the Gas Sampling System is connected properly.
To connect the Gas Sampling System, insert the Luer Lock connector to the Gas
Sampling input and rotate it clockwise one half turn. Now insert the Gas Sampling
elbow into the end of the Breathing Circuit.
7. Inspect the all ADS 2000 hoses periodically to make sure that they are properly
connected, and not kinked, cracked, or broken.