MAINTENANCE
Blanket Warmer Oper a tors Ma nu al M N- 39 9 2 4 Rev 4 8/20
40
Continued from previous page
Yearly For yearly maintenance, do the following.
Check:
the set-point temperature compared to the actual temperature displayed.
Check the cavity air temperature with a thermocouple placed 1” (25mm) from
the cavity sensor. Do not allow the thermocouple to touch any surface.
Monitor the temperature for approximately one hour in an empty cavity.
the condition of the plug and cord and replace if damaged.
the controller screen for excessive wear. Make sure the controller screen displays
and operates properly.
Clean:
the shelves and interior of the warmer. See topic How to Clean the Warmer for the
list of approved cleaners.