AlcoQuant® 6020 – Breathalyser – Operating Instructions
11
1111
11 Error description and rectification
Error description and rectificationError description and rectification
Error description and rectification
Error indication
Error indicationError indication
Error indication Possible cause
Possible causePossible cause
Possible cause Remedy
RemedyRemedy
Remedy
User error
User errorUser error
User error
Blow failure
Blow failureBlow failure
Blow failure
This error appears if during the
blowing-in process the respiratory
flow drops below the required level
before the blowing volume of at
least 1.2L was reached.
After the failure is output (approx. 2
seconds), the device switches back to the
previous measuring mode.
A signal tone and the flashing green LED
blinking at an interval of 1.5 seconds
signalise that the device is ready for
measuring.
The blowing process can be restarted.
The blowing process can be restarted. The blowing process can be restarted.
The blowing process can be restarted.
The test person should blow longer and
The test person should blow longer and The test person should blow longer and
The test person should blow longer and
harder than before.
harder than before.harder than before.
harder than before.
Device errors which do not endanger the meas
Device errors which do not endanger the measDevice errors which do not endanger the meas
Device errors which do not endanger the measuring process (yellow LED)
uring process (yellow LED)uring process (yellow LED)
uring process (yellow LED)
Signal generator
Signal generatorSignal generator
Signal generator
A defective signal generator has
been detected.
Measuring is still possible
The error message is only output during
the self test (while the device is being
switched on)
If the error recurs, please inform Service.
Error red LED (green, yellow)
Error red LED (green, yellow)Error red LED (green, yellow)
Error red LED (green, yellow)
A defective, or only weakly-lit LED
was detected.
Measuring is still possible
The error message is only output during
the self test (while the device is being
switched on)
If the error recurs, please inform Service.
Time
TimeTime
Time
An error was detected while the
current date/time was being
determined.
Measuring is still possible
The correct date/time may not be
displayed. If this is the case, incorrect date,
time values will be stored to memory with
the current measured value.
If the error recurs, please inform Service.