Cleaning and maintenance
48
Only touch new lamps in the handling zone.
Test the safety switch to ensure that it works.
The safety switch ensures that the device switches off in the event of a
faulty filter panel or when a filter panel is missing.
Press the button.
If the button springs back automatically then it is functional.
Replace the filter panel.
5.9 Clean and calibrate the sensor
Clean the sensor
NOTE:
Do not spray the sensor directly with detergent.
Always remove the sensor from the station for cleaning.
Remove the sensor from the base station.
Clean the sensor housing, measuring window and base station with a
soft, damp cloth. See separate instructions (order no. 1005902-..).
Replace the sensor into the base station.
WARNING!
Do not rub with a dry cloth – danger of scratching!
Failure to comply voids any warranty claims.
For the fast, hygienic and proper cleaning of the sensor
and base station, use specially developed cleaning
agents.
Concentrated disinfectants or solvents (e.g. Lysoform,
ethyl alcohol or other fluids containing alcohol) must
not be used.