Technical data
09/2011 1002410-04 Operating instructions – 65
AFFINITY 660 DYNAMIC POWER /
AFFINITY 660 DYNAMIC POWER IQ
Power consumption rating with air conditioning:
Power consumption rating without air conditioning:
13000 W
Rated frequency: 50 Hz
Rated voltage: 400-415 V 3N~ 230-240 V ~3
Rated fuses: 3 x 35 A (time delay) 3 x 50 A (time delay)
Connecting line: H05VV-F 5G 6 mm²
or
H05VV-F 4G 10 mm²
Noise level 1m away from device: 68.9 dB(A)
Noise level inside the device: <81 dB(A)
Load on the acrylic glass panel: max. 135 kg
Weight: ca. 515-584 kg