Captair Bio Smart
Mobile ductless ltering PCR workstations
16
Replacing the filters
Your device is equipped with FLEX™ lter technology that was congured to the user’s protection needs when the device
was purchased. The conguration of the lter column is dependent on the applications carried out in the enclosure. These
applications may change over time. Your FLEX™ lter technology can therefore be recongured if your workstation is used
for applications other than those anticipated when the device is rst set up. If so, please contact us so that we can verify that
the current conguration is safe or it needs to be recongured.
The table below summarises all possible Flex™ lter technology congurations for your device
Captair 321 - 391 - 481 Smart models
Molecular lter HEPA lter H14 / ULPA Pre-lter
Column
Congura-
tion
1C x1 x1
1C1P x1 x1 x1
UP
The table below summarises the different types of carbon lters that Erlab® offers along with their elds of application.
Type AS
For organic vapors
HEPA H14
For particles
ULPA U16
For particles