Start up
Calibrating the electronic anemometer
Open the sash of the workstation to perform calibration.
1- Press and hold the mute button located on the front of the control panel while simultaneously switching on the main power
button located behind the control panel (Off = O, On=I)
2- Release the mute button. A beep indicates that the unit has started the calibration mode. The white bar above the front of
the control panel will pulse for 5 minutes and 30 seconds. Stick with just the instructions for the anemometer.
3- Once the pulsing light stops, the anemometer is calibrated and the unit is ready to be turned on.
4- To operate, push the power button located on the front control panel.
Good practices :
Avoid any air turbulence in the area during calibration.
Anemometer should be calibrated every year and every time a modication is done on the ltration column.
We also recommend verifying the operating parameters before each new use.
To modify settings, please access the administrator interface.
Having carefully followed the steps described in the installation guide, your Captair Bio Smart ductless ltering PCR
workstation is now ready to use.