When reassembling the filtration column, be sure that the fan module is positioned correctly to access your connections points.
Air Flow
Reconnect the power supply to the fan
Recommendations for storing and using the filters
New lter shelf life and storage conditions:
New activated carbon molecular lters must be stored at in their original packaging at a temperature of between +10°C / 50°
F and +50°C / 120° F and a humidity level < 85% RH.
HEPA H14 and/or ULPA lters must be stored upright and kept dry.
If these conditions are adhered to, the maximum time a lter can be stored before use depends on the type of carbon used:
AS type carbon molecular lter 1 year from the date of delivery
HEPA/ULPA Filter 2 years from the date of delivery
If the maximum shelf life is reached, we recommend to not install lters.
We recommend replacing the lter annually (if used 24/7).
• Prelters(s) should be replaced every year.
• HEPA/ULPA lters should be replaced every 3 years on average.
• Molecular lter should be replaced every 2 years on average.
Replacement frequency of filtration breakthrough sensor,
Molecode option
Molecular lters saturation detector – Molecode S (solvents) is installed as an option in your unit and use a semi-conductor that
must be changed every 5 years.
After this period, Erlab cannot guarantee performances stability nor detection sensitivity.
Equipment replacement requires the intervention of a qualied engineer, able to replace internal components and perform
system new conguration. The change requires less than one hour.
Erlab Maintenance department and accredited partners can provide this service. (service not available in all countries).
For further information, please get in touch with your contact person or Erlab dealer/distributor or visit our website.