21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920
E-mail: ductless@escoglobal.com Website:
www.escoglobal.com/ductless
USER SECTION
CHAPTER 4 – OPERATING YOUR CABINET
4.5 CLEANING AND SHUTTING DOWN THE CABINET
1. As far as possible, it is recommended that the cabinet be operated continuously in order to
achieve optimal containment and cleanliness.
NB: Airflow studies have shown that once the fan has been switched off, air from the cabinet may
escape due to the thermal currents from inside the cabinet. So if you choose to turn off the cabinet,
you need to carefully adhere to the following guidelines so as NOT to compromise safety and
containment during shutdown.
2. Thoroughly wipe the work surface, side walls, inner back wall and drain pan(not applicable for SPB)
using water and some mild detergent. Periodically clean behind the sash window too.
3. Wipe off the detergent using water. Make sure that there are no traces of detergent left.
4. Leave the blower on for 3 minutes in order to purge the work zone.