Polymerase Chain Reaction Cabinets
Fig 2.3
Blower does not
function
Faulty motor speed
controller
Unplug the cabinet from main supply.
Open electrical panel and see Layout A/B at the end of
this section to locate the motor speed controller.
Add a jumper wire between the 2 cables going to speed
control (see Fig. 2.4, 2.5).
Plug the cabinet to the main supply.
If the fan operates properly, the speed controller is
defective and need to be replaced.
If the fan does not operate, the problem is not with the
speed control. Proceed to next step.
Fig 2.4
Fig 2.5