Esco SCR-2A2 - Problem 3 - Airflow is not at nominal velocity

Esco SCR-2A2
63 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
29
24
Blower does not
function
Auto-thermal cut-off
Check the Fan for overheating The motor has an in-
built thermal cut off.
Wait 60 minutes with the FAN turned off and then try to
restart.
If the FAN restarts determine why there is excessive
heat in the cabinet.
If the FAN does not restart proceed to the next step.
Motor failure
Unplug the cabinet from the main supply.
Check motor condition. If physical damage is present, like
fan blades bent, cracked or chipped off, replace blower
(see Fig. 2.9).
To replace the blower, please follow instruction which is
available in this manual under Service Section.
Fig 2.9