9.5.6 Setting alarm parameters
Item
Adult Range
Child Range
Accuracy
MV-upper limit
099L
120L
0.5L or ±15%
MV-lower limit
098L
019L
0.5L or ±15%
P
aw
-upper limit
1 cmH
2
O 80
cmH
2
O
1 cmH
2
O40 cmH
2
O
±2 cmH
2
O or ±10
P
aw
-lower limit
079 cmH
2
O
039 cmH
2
O
±2 cmH
2
O or ±10
V
TE
-upper limit
50ml2000ml,
50ml600ml
±20mL or ±15%
V
TE
-lower limit
OFF,401990ml
OFF,40ml590ml
±20mL or ±15%
Rate-upper limit
1100 bpm
170 bpm
2 bpm020bpm);
10%other
Rate-low limit
099bpm
069bpm
2 bpm020bpm);
10%other
Tapnea
15s60s
15s60s
±1s or ±10%
FiO
2
-lower limit
21%~99%
21%~99%
±3%
FiO
2
-upper limit
22%~100%
22%~100%
±3%
CAUTION: All low limits of parameters in above table may not be set up the high
limits, nor may the high limits be set below the low limits.
9.6 Order information
Description
specification
Stock number
Exhalation valve diaphragm
SH300-30-03
Exhalation valve
SH300-30-00
One-way membrane
SH300-30-05
Filter
VS-21-01
O
2
pipeline
5m
GYG-01A
Air pipeline
5m
GYG-03B
Power cord
3m
ES-03
O
2
sensor
---
EK-12
Silicon corrugated hose
---
ACC-27
Manual gasbag
3L
ACC-02
Cleat gasbag
---
ACC-05
Face mark
---
ACC-04