9.5.6 Setting alarm parameters
Item
Adult Range
Child Range
Accuracy
MV-upper limit
0~99L
1~20L
0.5L or ±15%
MV-lower limit
0~98L
0~19L
0.5L or ±15%
P
aw
-upper limit
1 cmH
2
O ~ 80
cmH
2
O
1 cmH
2
O~40 cmH
2
O
±2 cmH
2
O or ±10%
P
aw
-lower limit
0~79 cmH
2
O
0~39 cmH
2
O
±2 cmH
2
O or ±10%
V
TE
-upper limit
50ml~2000ml,
50ml~600ml
±20mL or ±15%
V
TE
-lower limit
OFF,40~1990ml
OFF,40ml~590ml
±20mL or ±15%
Rate-upper limit
1~100 bpm
1~70 bpm
2 bpm(0~20bpm);
10%(other)
Rate-low limit
0~99bpm
0~69bpm
2 bpm(0~20bpm);
10%(other)
Tapnea
15s~60s
15s~60s
±1s or ±10%
FiO
2
-lower limit
21%~99%
21%~99%
±3%
FiO
2
-upper limit
22%~100%
22%~100%
±3%
CAUTION: All low limits of parameters in above table may not be set up the high
limits, nor may the high limits be set below the low limits.
9.6 Order information
Description
specification
Stock number
Exhalation valve diaphragm
SH300-30-03
Exhalation valve
SH300-30-00
One-way membrane
SH300-30-05
Filter
VS-21-01
O
2
pipeline
5m
GYG-01A
Air pipeline
5m
GYG-03B
Power cord
3m
ES-03
O
2
sensor
---
EK-12
Silicon corrugated hose
---
ACC-27
Manual gasbag
3L
ACC-02
Cleat gasbag
---
ACC-05
Face mark
---
ACC-04