9.7 Electromagnetic Compatibility
Changing or reassembling this equipment without Eternity‟s authorization may cause
electromagnetic compatibility problems. Contact with Eternity for assistance. Designing and
testing this equipment is in accordance with the following stipulations.
WARNING: using cell phone or other radio radiant equipment near this product
may cause malfunction. Closely monitor the working condition of this
equipment if there is any radio radiant supply nearby.
Using other electrical equipment in this system or nearby may cause interference. Check if the
equipment works normally in these conditions before using on a patient.
Be careful of the following when SH300 is connected:
Do not put any object which is not in accordance with EN60601-1 in the 1.5M range of patients.
An isolated transformer must be used for alternating current supply (in accordance with IEC60989),
or additional protective ground wires are equipped if all the devices (for medical or non-medical
use) are connected to SH300by using signal input/signal output cable.
If a portable all-purpose outlet is used as the alternating current supply, it must be in accordance
with EN60601-1-1 and cannot be put on the floor. Using another portable all-purpose outlet is not
recommended.
Do not connect the non-medical equipment directly to the alternating current outlet on the wall.
Only the alternating current supply of the isolated transformer can be used. Otherwise, the surface
leaking current may exceed the range permitted by EN60601-1 under the normal conditions, and
misoperation may cause injury to patients or operators.
A complete system current leaking test (according to EN60601-1) must be performed after any
equipment is connected to these outlets.
WARNING: medical electrical equipment operators contact non-medical electrical
equipment and patients at same time. It is dangerous of patients or
operators.