60
Guidance and manufacture’s declaration electromagnetic emissions-
for all EQUIPMENT and SYSTEMS
Guidance and manufacture’s declaration – electromagnetic emission
The SH300 Ventilator is intended for use in the electromagnetic environment specified below. The
customer or the user of SH300 Ventilator should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment guidance
RF emissions
CISPR 11
Group 1
The SH300 Ventilator uses RF energy only for
its internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
RF emission
CISPR 11
Class B
The SH300 Ventilator is suitable for use in all
establishments, including domestic and those
directly connected to the public low-voltage
power supply network that supplies buildings
used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Guidance and manufacture’s declaration electromagnetic immunity
for all EQUIPMENT and SYSTEMS
Guidance and manufacture’s declaration – electromagnetic immunity
The SH300 Ventilator is intended for use in the electromagnetic environment specified
below. The customer or the user of SH300 Ventilator should assure that it is used in such an
environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic
environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
6 kV contact
8 kV air
6 kV contact
8 kV air
Floors should be wood,
concrete or ceramic
tile. If floors are
covered with synthetic
material, the relative
humidity should be at
least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
2 kV for power
supply lines
1 kV for
input/output lines
2kV for power
supply lines
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC 61000-4-5
1 kV differential
mode
2 kV common
1 kV differential
mode
2 kV common
Mains power quality
should be that of a
typical commercial or