SYSTEM PARAMETER SETUP
33
8.6 Single Step
Press the SETUP key to advance the menu to the Single step parameter. The
display will show:
Single Step = ON
ENTER to select
or Single Step = OFF
ENTER to select
Use the ENTER key to select the Single step mode.
After the selection of Single step mode, press the SETUP key to advance to the
Memory Lock selection, or press the EXIT key to exit from the Single step
selection to the operation mode. The Program will store the selection of Single step
automatically.
This function is used to temporarily override the automatic connection feature.
When the single step function is ON the instrument will pause after each step is
completed. To continue the test sequence, press the Test button to execute the next
connected step. Each time the Test button is pressed the next connected step will
execute. The first step of the sequence will execute, when the Test button is
pressed after the last step of the sequence has completed.
8.7 Memory Lock selection
Press the SETUP key, the program will advance to the Memory Lock selection
mode. The display will show:
Memory Lock = O N
ENTER to select
or Memory Lock = O F F
ENTER to select
If the Memory Lock is selected ON, the Memory selection Menu will be disabled
when in Lock Mode. If the Memory Lock is selected OFF, the Memory selection
Menu will be enabled when in Lock Mode. Different memories can be recalled but
the test parameters or steps can not be changed. The Memory Lock default is
preset to ON at the factory.
After the selection of Memory Lock mode, press the SETUP key to advance to the
beginning of the menu (PLC Remote), or press the EXIT key to exit from the
Memory Lock selection to the operation mode. The program will store the selection
of Memory Lock automatically.