LCD DISPLAYED MESSAGES
37
CHAPTER 10 : LCD Displayed Messages
10.1 Test In Process
After pressing the test button to initialize the test, the TEST light on the right side
of the instrument will illuminate. The instrument will indicate the test is
processing as follows:
Test 222s 110.2V
M1-1
_
G-L
2500
µ
A
10.2 Abort
If the RESET button is pressed during the test cycle the test will be aborted and
the display indicate that the test has been aborted as follows:
A b o r t D 110.2V
M1-1
_
G-L
2500µ A
10.3 Pass
When the test cycle has completed successfully without detecting any failures, you
will hear a short audible beep to let you know the test has completed successfully.
The display will also indicate that the test has passed as follows:
P a s s D 110.2V
M1-1
_
G-L
2500
µ
A
10.4 High Limit Failure
If the leakage current of the item under test exceeds the HI-Trip setting, the unit
will indicate a failure at the end of the delay time by illuminating the FAIL
indicator and activating the alarm. All relays will be deactivated to disable
voltage to the DUT. The display will indicate a high limit failure as follows:
H I - F a i l D 110.2V
M1-1
_
G-L
2500µ A
To clear the failure light and alarm, press the RESET button. The alarm and light
will be turned off but the display will continue to indicate the type of failure until
a second reset is activated.