Datalogger Mode6
2 Language Set the report language to English, Ger-
man, French, Italian, Spanish, or
Portuguese.
English
3
Sampling
Rate
The rate at which readings are logged.
Select from 30 seconds to 2 hours.
5 minutes
4
Start Delay Set a delay so that the logger will not start,
after the Start button is pressed, until the
delay time elapses. Can be set up to 24
hours.
0 minutes
5
Unit of
Measure
Select ℃ or ℉ temperature units. ℃
6 Password
Set a 16 character (maximum) alpha-nu-
meric password.
Disable
7
Company
Name
Add a 20 character (maximum) text string
to logging reports.
n/a
8
SAVE Click SAVE to complete the configuration. n/a
9 Alarm Types
Single Alarm: Triggers immediately when
measurement exceeds the alarm limit.
Cumulative Alarm: Triggers when the
average measurement exceeds the limit.
Disable Alarm: No Alarm. See Section 7
for complete Alarm information.
Disable
10 Alarm Delay For Cumulative alarms only, a delay inter-
val can be set to delay an alarm trigger
until after the delay timer elapses. Adjust-
able from 5 minutes to 2 hours.
0 minutes
11 Temp. Alarm
Limit
Set the Temperature Alarm threshold
range from -22 to 158℉ (–30 to 70℃).
32 to 140℉ (0 to
60℃)
12 RH Alarm
Limit
Set the RH Alarm threshold range from 0
to 100% RH.
10 to 90%
13
Cancel Click to abort a configuration session. n/a
To complete and confirm the configuration, click ‘SAVE’, close the setup win-
dow, eject the 42280A from the PC (right-click the drive icon on the PC and
select ‘Eject’), and then disconnect the logger from the USB port.
Logging can now be started by long pressing the START/STOP button until
REC appears on the display.