Programming Parameters - Conductivity
The Table below shows the available programming menu items P1 ~ P7. Each parameter
is explained in detail in the subsequent sections.
Parameters Code Selections
P1 Standard solution series selection
USA (Europe & U.S.A)
CH (China)
P2
Electrode Constant selection
0.1, 1, or 10
P3
Reference Temperature selection
77, 68, and 64
o
F (25,
20,and 18 ℃)
P4 Temperature compensation
coefficient setting
0.00 to 9.99%
P5 Electrode Constant Calibration
P6 Temperature units ℃ ℉
P7 Back light display time setting 0-1-3-6 minutes
P8 Auto power off setting 0-10-20 minutes
P9 Restore to default factory settings OFF / ON
Parameter P1 (pH Buffer Solution Setting)
1. From the pH measurement mode, press and hold MODE for at least 2 seconds and
then release, the ‘P1’ icon appears on the LCD
2. Use the CAL or the M+/RM keys to toggle through the two (2) selections: USA (for use
in the USA or Europe) and CH (for use in China)
3. Momentarily press MODE to move to the next parameter (P2), or press ENTER to
return to the normal measurement mode
Parameter P2 (Electrode Constant ‘K’ Selection)
1. From the P2 menu, use the CAL or the M+/RM keys to select the desired electrode
constant (0.1, 1.0, or 10). The default setting is K=1.0
2. Momentarily press MODE to move to the next parameter (P3), or press ENTER to
return to the normal measurement mode