6
SDL600-EU-EN-V1.2-10/11
Enabling/Disabling the Auto Power OFF Feature
1. Access the PoFF parameter.
2. Use the arrow buttons to select ON (enable) or OFF (disable). With the Auto Power OFF
feature enabled, the meter will automatically switch OFF after 10 minutes of inactivity.
3. Press ENTER to confirm setting.
4. Press and hold the SET button for at least 1.5 seconds to exit to the normal operation mode (or
simply wait 7 seconds for the meter to automatically switch to the normal operating mode).
Set the Beeper Sound ON or OFF
1. Access the bEEP parameter.
2. Use the arrow buttons to select ON (enable) or OFF (disable).
3. Press ENTER to confirm setting.
4. Press and hold the SET button for at least 1.5 seconds to exit to the normal operation mode (or
simply wait 7 seconds for the meter to automatically switch to the normal operating mode).
Numerical Format (comma or decimal)
European and USA numerical formats differ. The meter defaults to USA mode where a decimal
point is used to separate units from tenths, i.e. 20.00; The European format uses a comma, i.e.
20,00 to separate units from tenths. To change this setting:
1. Access the dEC parameter.
2. Use the arrow buttons to select USA or Euro.
3. Press ENTER to confirm setting.
4. Press and hold the SET button for at least 1.5 seconds to exit to the normal operation mode (or
simply wait 7 seconds for the meter to automatically switch to the normal operating mode).
SD Card FORMATTING
1. Access the Sd-F parameter.
2. Use the arrow buttons to select YES to format the card (select NO to abort). Note that all data
on the card will be lost if formatting is attempted.
3. Press ENTER to confirm selection.
4. Press ENTER again to re-confirm.
5. The meter will automatically return to the normal operating mode when formatting is complete.
If not, press and hold the SET button for at least 1.5 seconds to exit to the normal operation
mode.
Set the Frequency Response type (‘A’ or ‘C’ Weighting)
1. Access the SET parameter.
2. Use the arrow keys to select A or C.
3. Press ENTER to confirm setting.
4. Press and hold the SET key for at least 1.5 seconds to exit to the normal operation mode (or
simply wait 7 seconds for the meter to automatically switch to the normal operating mode).

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Extech Instruments SDL600 and is the answer not in the manual?

Extech Instruments SDL600 Specifications

Display IconDisplay
display size52 x 38mm (2 x 1.5”) Backlit LCD
display update rateApprox. 1 second
Microphone IconMicrophone
microphone type12.7mm (0.5”) electret condenser microphone
microphone diameter12.7mm (0.5”) outer diameter
Measurement IconMeasurement
measurement unitsDecibels
measurement ranges30 to 80, 50 to 100, and 80 to 130dB
frequency range31.5Hz to 8KHz
Power IconPower
power supplySix (6) 1.5 VDC batteries (optional 9V AC adaptor)
power consumption (normal operation)approx. 12mAdc
power consumption (backlight off, datalogging on)approx. 51mAdc
Operating Environment IconOperating Environment
operating temperature0 to 50°C (32 to 122°F)
operating humidity85% R.H. max.
auto power offAfter 10 minutes of inactivity (can be disabled)
Weight and Dimensions IconWeight and Dimensions
weight428g (0.94 lbs.) including batteries
main instrument dimensions250 x 73 x 47.5mm (9.8 x 2.9 x 1.9”)