a D:GROUP company
00020EN Rev.09 – 12/2012 19
5 LIMITATIONS
PRECAUTIONS FOR THE CORRECT USE OF THE CABINET.
Listed below are the most important guidelines to be followed and the main substances to be avoided to ensure the
correct use of the FLOWFAST H cabinet:
-
NEVER handle pathogenic materials of any group or biosafety level.
-
NEVER USE chlorine-based substances (e.g. sodium hypochlorite) as they are corrosive for the metal structure
of the cabinet, and in particular for stainless steel parts.
-
DO NOT use ethanol as a sterilizing substance if a heat source is used under the cabinet.
-
DO NOT use cosmetic powders, nail polish, hairspray or cosmetics in general during work.
-
DO NOT eat, drink or smoke in the work zone.
-
AVOID substances that release explosive vapours.
In addition, when working with the cabinet, AVOID:
-
the introduction of extraneous material
-
quick movements of arms
-
upstream contamination of the material, putting the hands or any object between the absolute filter and the
sterile material
-
working under the cabinet if the airflow has not been activated yet, and a suitable work condition has not been
reached yet. After the cabinet is switched on, and if the work requires special sterile conditions, chemical
sterilization of the work chamber must be carried out using a cloth soaked in bactericide. Then wait for 20-30
minutes for the bactericide to take effect.
IMPORTANT: The UV radiations emitted by the germicidal UV lamp can cause erythemas and
conjunctivitis. Avoid the exposure of skin and eyes to direct radiation.