8 TROUBLESHOOTING PROBABLE CAUSES OF MALFUNCTIONS
PROBLEM CAUSE REMEDY
Cabinet does not work the electricity supply has
been cut off at the mains
Electronic board out of order
Blown fuse(s)
check the voltage input to the
cabinet
Replace PCB
Replace fuses
Alarm: “HEPA filters check” Main HEPA filter clogged Replace HEPA filter
HEPA filters clogged . Replace HEPA filters. “minimum LAF alarm”.
[Possible contamination of the
product but protection of the
environment]
The main motor-fan does not work
Check the terminal voltage of the
power card of the main motor-fan
Check F7 fuse on the power card
Replace the power card
Replace the microprocessor card
“Black-out ” Alarm
(probable exchange of air between
the work chamber and the outside
and possible contamination of the
environment)
Blackout
Check the feeding cable, the
connection plug/socket, the power
supply line
Press “ESC” [7] key to silence the
alarm
Alarm “Sensors failure” Failure of sensor XX Replace XX sensor.
Alarm: “No encoder input LAF” No signal from LAF flow sensor Replace the LAF sensor