17.6 Detection of Long Measurement Failures
If there are no valid measured value during a long time interval, new increments of the totalizers will be ignored. The val-
ues of the totalizers remain unchanged.
In the SuperUser mode, it is possible to set the time interval. Select Special Funct.\SYSTEM settings\Measur-
ing\Miscellaneous. Press ENTER until the menu item Do not total. if no meas. is displayed.
17.7 Number of Decimal Places of the Totalizers
The value of a totalizer consists of max. 11 digits, including max. 4 decimal places. In the SuperUser mode, it is possible
to define the number of decimal places.
Select Special Funct.\SYSTEM settings\Measuring\Miscellaneous. Press ENTER until the menu item To-
tal digits is displayed.
Total digits = Automatic
The number of decimal places will be adjusted dynamically. Low totalizer values will first be displayed with 3 decimal plac-
es. For higher totalizer values, the number of decimal places will be reduced.
Total digits = Fixed to x digit
The number of decimal points is constant. The max value of the totalizer is reduced with the number of decimal places.
For setting the behavior of the totalizers after reaching the max. value see section 13.3.1.
17.8 Manual Reset of the Totalizers
If the manual reset of the totalizers is activated, the totalizers can be reset to zero during the measurement by pressing
key CLR three times, even if a program code is activated.
Select Special Funct.\SYSTEM settings\Measuring\Miscellaneous. Press ENTER until the menu item 3xC-
clear totals is displayed.
Enter the time. If 0 (zero) is entered, the default value 30 s will be used.
Select one of the following list items.
Automatic: dynamic adjustment
Fixed to x digit: x decimal places (range: 0...4)
Press ENTER.
max. value display
< 10
6
±0.000 ... ±999999.999
< 10
7
±1000000.00 ... ±9999999.99
< 10
8
±10000000.0 ... ±99999999.9
< 10
10
±1000000000 ... ±9999999999
decimal places max. value max. display
0< 10
10
±9999999999
1< 10
8
±99999999.9
2< 10
7
±9999999.99
3< 10
6
±999999.999
4< 10
5
±99999.9999
Note! The number of decimal places and the max. value defined here only affect the display of the totalizers.
Select on to activate the manual reset of the totalizers, off to deactivate it. Press ENTER.
Note! The manual reset of the totalizers is still active after the deactivation of the SuperUser mode.
Do not total. if
no meas.> 0 s
Total digits
Automatic
3xC clear totals
off >ON<

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Flexim FLUXUS G800 and is the answer not in the manual?

Flexim FLUXUS G800 Specifications

General IconGeneral
BrandFlexim
ModelFLUXUS G800
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals