FLUXUS G80x 15 Libraries
110 UMFLUXUS_G8V4-6-1EN, 2018-10-10
Material Parameter
Enter the material's:
• transversal sound speed
• longitudinal sound speed
1...5 values depending on the selected function have to be entered. Press ENTER after each input.
If an already defined material is edited, for each parameter there will be a request whether it is to be edited.
Select yes or no. Press ENTER. Change the values, if necessary.
Fluid Parameter
Enter the fluid's:
• longitudinal sound speed
• kinematic viscosity
•density
• gas compressibility coefficient
Depending on the selected function, 1...5 values have to be entered. Press ENTER after each input.
If an already defined fluid is edited, for each parameter of some of the functions there will be a request whether it is to be
edited. Select yes or no. Press ENTER. Change the values, if necessary.
15.4 Deleting a User Defined Material/Fluids
To delete a user defined material/fluid, proceed as follows:
Select Special Funct.\Install Material or Install Medium. Press ENTER.
If the extended library is activated, press ENTER until the request for deleting is displayed.
Select the type of sound wave to be used for the flow measurement. Press ENTER.
For most materials, a transversal sound wave has to be selected.
Enter the typical roughness of the material. Press ENTER.
Select yes to store the entered parameters or no to quit the menu item without storing.
Press ENTER.
Select yes to store the entered properties, no to quit the menu item without storing. Press
ENTER.
Select delete. Press ENTER.
Select the material/fluid to be deleted. Press ENTER.
Select yes or no. Press ENTER.
Default soundsp.
long.   >TRANS.<
Install Material
edit    >DELETE<
USER Material
#01: Polystyrene