Handling the camera11
11.19 Calibrating the lens–camera combination
Applicability: Camera models with an exchangeable lens.
11.19.1 Introduction
Before a new lens can be used with the camera, the lens–camera combination must be
calibrated.
This is a process that previously had to be performed by a FLIR service department, but
for the FLIR Exx series the calibration can be performed by the user. This feature is called
AutoCal. The AutoCal procedure requires a calibration target, which is included in the lens
package.
11.19.2 AutoCal procedure
Follow this procedure:
1. Dip the calibration target in water for 1 second and let the excess drip off.
#T810190; r. AL/47698/47698; en-US