EasyManua.ls Logo

FLIR MR277 - Safety Specifications; MR13 Sensor Measurement Specifications

FLIR MR277
72 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Specifications
14
14.5 Safety Specifications
General Safety CE/EN/UL/CSA/PSE 61010
Environmental Safety REACH Regulation EC 1907/2006
RoHS2 Directive 2011/65/EC
WEEE Directive 2012/19/EC
JIS C 6802:2011 laser directive
IEC 60825-1 class II laser directive
FDA Laser directive
Encapsulation IP54 (IEC 60529) with bottom flap fully sealed
Drop-proof Designed to 6.6 ft. (2 m)
EMC EN 61000-6-3
EN 61000-6-2
FCC 47 CFR Part 15 Class B
14.6 MR13 Sensor Measurement and Calculation Specifications
Parameter Range Accuracy Resolution
Relative Humidity 10-90% ±2.5% 0.1
Air Temperature 32 to 122
(0 to 50)
±1.1 (±0.6)
Dew Point -22 to 122
(-30 to 50)
n/a (calculation)
Vapor Pressure 0.0–12.0 kPa
n/a (calculation)
Mixing Ratio
0-560 GPP
(0.0-80.0 g/kg)
n/a (calculation)
NOTE
Relative Humidity Sensor Note:
The meter should be stored in an environment with a relative humidity similar to the area
to be tested. If the storage humidity differs by more than 50% RH from the area under test,
an acclimation period of up to 24 hours may be required to meet the specified RH
accuracy.
#NAS100005; r. AG/62027/62027; en-US