FLIR MR55 USER MANUAL Document Identifier: MR55-en-US_AA
9
4.10 Calibration Verification Test
With the meter ON and set to Group 1, touch the measurement pins to the test
points labeled ‘T’ on the protective cap (test points ‘1’ in accompanying diagram).
The display will read 18% (±2) if the test passes. Next, touch the measurement pins
to the test points labeled ‘B’ on the protective cap (test points ‘2’ in the
accompanying diagram). The display will read 26% (±2) if the test passes. Please
return the MR55 for service if either self-test consistently fails.
5. Bluetooth®
Technical Details
When connected with a remote device running the FLIR Tools mobile app, the MR55
(using the METERLiNK® protocol) can send readings for display on the remote device.
When connected to a compatible FLIR camera that supports BLE (Bluetooth® Low
Energy), the MR55 can send meter readings for display on the camera screen.
Download the FLIR Tools mobile app from the Google Play Store, the Apple App
store, or from here: https://www.flir.com/products/flir-tools-app/
1. When successful communication between the meter and a remote device or FLIR
camera is established, the Bluetooth® icon appears on the meter display.
2. Refer to the FLIR Tools help utility (in the mobile app) for detailed information
and tutorials regarding the FLIR Tools application.
3. Refer to Section 4.4 Bluetooth® Transmission Basics for using the meter to
transmit data via Bluetooth®.