9. Diagnostics
If your camera does not function exactly as it should, check the following table
before consulting FLIR after-sales service or your FLIR representative.
9.1. Diagnostic table
Surface treatment of lens
is scaled
Performances of optics will be degraded.
Metrological measurements are not guaranteed
any more (Altaïr - image in °C)
Contact FLIR After Sale Service
The cooler does not make
any more noise
First of all, check that the camera is properly
connected and powered. Connections are
correct?
No remake connections
Yes Contact FLIR After Sale Service
The Stirling engine does
not seem not to reach its
specified point
It always turns to full power. Check through
CIRRUS software the temperature of the
detector after 10 min of powering. It is close to
temperature specified in the data sheet (+/- 2
K)?
Yes The temperature control is done
correctly, but may be overworking the
Stirling engine. Stay alert to any change
that may require a preventive intervention.
No Contact FLIR After Sales Service
Difficulty installing a
connector
Pins are twisted?
No Engage the connector gently.
Yes Envisage the replacement of the cable.
Contact FLIR After Sales Service
Inconsistency of the IR
image in Altaïr
Have you deactivated all process Plugin (Filter,
threshold...)?
No Deactivate them. Is the image still
inconsistant ?
Yes Check the cable and the connectors.
No The problem is solved.