[5.5.2.4 — Area]
ThermaCAM™ PM575/595 Operator’s Manual
32
button to resume the normal function of the joystick, i.e. focus/zoom. To move the
crosshair next time, it is necessary to confirm Spot (or Spot 1, 2 or 3) in the Anal-
ysis menu.
In Maximum (or Minimum) mode four corners of a box are displayed on the screen
when the spotmeter function is activated. The size of the box can be controlled by
using the joystick. When the desired box size has been set press the
button to
resume the normal function of the joystick. To change the size of the box next time,
it is necessary to confirm Spot again (or Spot 1 if Maximum or Minimum is
selected in the Setup – Analysis dialog box; Setup
Analysis).
The crosshair will now continuously move to the part of the image within the box
where it hits the maximum (or minimum) temperature.
5.5.2.4 Area
The area function measures either the maximum, minimum or average temperature
within a rectangular or circular area around the centre of the image. The result is
indicated by the Amin, Amax or Aavg text in the result table. To resize the rectan-
gle, confirm Area in the Analysis menu. The size of the area can then be controlled
by the movement of the joystick in the following way while the text Joystick: Area
size C: Remove is shown at the bottom of the screen. To change the size of the
area next time, it is necessary to confirm Area in the Analysis menu again.
Joystick Up: Increases the rectangle height or circle radius.
Joystick Down: Decreases the rectangle height or circle radius.
Joystick Left: Decreases the rectangle width or circle radius.
Joystick Right: Increases the rectangle width or circle radius.
Once the required area size has been set, press the
button to resume the nor-
mal function of the joystick, i.e. focus/zoom.
The Area type (i.e. Box or Circle) and the Area function (i.e. Amin, Amax or Aavg)
are selected in the Setup
AnalysisArea type and SetupAnalysisArea
function menu (see “Setup menu” on page 35).
5.5.2.5 Profile
The profile function shows a horizontal or vertical temperature graph, depending
on the Setup
AnalysisProfile setting (see “Setup menu” on page 35), marked
by the profile line. The temperature of the cursor is indicated by the Cur text in the
result table. To move the cursor and the line along which the measuring takes
place, confirm Profile in the Analysis menu. The position of the cursor can then be
controlled by the movement of the joystick while the text Joystick: Prof pos C:
Remove is shown in the bottom left-hand corner of the viewfinder. Once the

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the FLIR ThermaCAM PM575 and is the answer not in the manual?

FLIR ThermaCAM PM575 Specifications

General IconGeneral
BrandFLIR
ModelThermaCAM PM575
CategoryThermal cameras
LanguageEnglish

Related product manuals